果糖胺
比色法
色谱法
化学
医学
内科学
胰岛素
作者
J R Baker,P A Metcalf,R N Johnson,David J. Newman,P Rietz
出处
期刊:Clinical Chemistry
[American Association for Clinical Chemistry]
日期:1985-09-01
卷期号:31 (9): 1550-1554
被引量:189
标识
DOI:10.1093/clinchem/31.9.1550
摘要
We have developed an automated colorimetric assay for glycated serum proteins (or fructosamines), measuring the reducing activity of serum in alkaline solution (pH 10.35) at 37 degrees C. The calibrants were prepared from a synthetic fructosamine (1-deoxy-1-morpholinofructose), although secondary standards of glycated bovine albumin were more robust in routine application. Interference was appreciable only with icteric specimens (bilirubin greater than 60 mumol/L), and between-batch imprecision (CV) was less than 2%. The range of fructosamine concentrations measured in 502 healthy (nondiabetic) blood donors was 1.87-2.87 mmol/L. There were no significant (p greater than 0.05) age- or sex-related differences in this population sample. Fructosamine accurately reflected blood glucose control as evidenced by the significant correlation with glucose concentrations in fasting plasma (r = 0.82, p less than 0.001) and with glycated hemoglobin (HbA1c) (r = 0.87, p less than 0.01) in 115 patients with type 2 (non-insulin-dependent) diabetes mellitus. The test is simple and rapid to perform (75 samples per hour) and provides an alternative to HbA1c determinations for monitoring blood glucose control and assessing the effects of changes in diabetes management.