心脏骨骼
医学
血流动力学
主动脉瓣置换术
假肢
心脏病学
主动脉瓣
内科学
外科
狭窄
作者
Giordano Tasca,Antonello Stefano Martino,Floriana Giannico,Beatrice Riva,Paola Redaelli,Elisabetta Lobiati,Michele Triggiani,Andrea Galanti,Amando Gamba
出处
期刊:PubMed
日期:2015-05-01
卷期号:24 (3): 360-7
被引量:4
摘要
Aortic valve replacement in patients with a small aortic root may be associated to high residual gradients. In such patients, both stentless valves and aortic annulus enlargement can reduce these residual gradients. Several studies have reported that Trifecta valves yield very good hemodynamic results. The aim of the present study was to compare the hemodynamic performance of Trifecta vs. Freestyle valves at one year in patients with an aortic annulus ≤ 2.3 cm.Between September 2011 and September 2013, 40 patients with a native aortic annulus diameter ≤ 2.3 cm and average age of 81 ± 4 years, were randomized to receive either a St-Jude Trifecta stented prosthesis (20 patients) or a Medtronic Freestyle stentless prosthesis (20 patients).No differences between Trifecta and Freestyle were found at one year in mean gradient s: 6.1 ± 3 mmHg and 6.6 ± 3 mmHg (p = 0.796); effective ori fice area: 1.82 ± 0.3 mmHg and 1.76 ± 0.4 mmHg (p = 0.676) or regression of left ventricular mass: - 25% ± 14 vs. -19% ± 16 (p = 0.204), respectively. Only moderate patient -pro sthesis mismatch was found, which affected 3 patient s in each group.At one year both stentless and stented prostheses yielded comparable hemodynamic results. These data suggest that Trifecta implantation is a valid means of avoiding patient -prosthesis mismatch in aortic valve replacement in elderly patients with a small native aortic annulus.