Laura Remón,Iulen Cabeza‐Gil,Begoña Calvo,Francisco Poyales,Nuria Garzón
标识
DOI:10.3928/1081597x-20200713-02
摘要
PURPOSE: To assess the biomechanical stability of three different marketed intraocular lenses (IOLs) with different haptic designs (four-loop IOL [Micro F FineVision model] and double C-loop IOL [POD F and POD FT models], all manufactured by PhysIOL), in silico (computer simulation) and in vivo (in the context of lens surgery). METHODS: An in silico simulation investigation was performed using finite element modeling (FEM) software to reproduce the compression test defined by the International Organization for Standardization and in vivo implantation in patients in the context of lens surgery was evaluated 1 day and 3 months postoperatively. IOL decentration and rotation were tested. In addition, the stress and strains were analyzed with the finite element method. RESULTS: In the in silico evaluation, the compression force for the POD F IOL was slightly lower than for the POD FT IOL and Micro F IOL for all compression diameters. The axial displacement was maximum for the POD FT IOL and the tilt, rotation, and lateral decentration were substantially lower than the acceptable tolerance limits established in ISO 11979-2. In the in vivo evaluation, a total of 45 eyes from 45 patients were selected, 15 eyes for each IOL model under assessment. Statistically significant differences were found between the Micro F and POD F IOLs for lateral decentration in x-direction (in absolute value) at 3 months postoperatively (P = .03). CONCLUSIONS: Although statistically significant differences have been found when comparing the displacement, tilt, and rotation between the different lenses, these differences cannot be considered clinically relevant, which would suggest that all three IOL models yield excellent stability in those terms. FEM appears to be a powerful tool for numerical studies of the biomechanical properties of IOLs. [J Refract Surg. 2020;36(9):617-624.].
科研通智能强力驱动
Strongly Powered by AbleSci AI