沙林
化学
离子迁移光谱法
塔邦
神经毒剂
索曼
质谱法
离子
化学战剂
分光计
电离
分析化学(期刊)
色谱法
有机化学
酶
生化工程
工程类
物理
乙酰胆碱酯酶
量子力学
作者
André Ahrens,Maria Allers,Henrike Bock,Moritz Hitzemann,Arne Ficks,Stefan Zimmermann
出处
期刊:Analytical Chemistry
[American Chemical Society]
日期:2022-10-27
卷期号:94 (44): 15440-15447
被引量:20
标识
DOI:10.1021/acs.analchem.2c03422
摘要
A growing demand for low-cost gas sensors capable of detecting the smallest amounts of highly toxic substances in air, including chemical warfare agents (CWAs) and toxic industrial chemicals (TICs), has emerged in recent years. Ion mobility spectrometers (IMS) are particularly suitable for this application due to their high sensitivity and fast response times. In view of the preferred mobile use of such devices, miniaturized ion drift tubes are required as the core of IMS-based lightweight, low-cost, hand-held gas detectors. Thus, we evaluate the suitability of a miniaturized ion mobility spectrometer featuring an ion drift tube length of just 40 mm and a high resolving power of Rp = 60 for the detection of various CWAs, such as nerve agents sarin (GB), tabun (GA), soman (GD), and cyclosarin (GF), as well as the blister agent sulfur mustard (HD), the blood agent hydrogen cyanide (AC) and the choking agent chlorine (CL). We report on the limits of detection reaching minimum concentration levels of, for instance, 29 pptv for sarin (GB) within an averaging time of only 1 s. Furthermore, we investigate the effects of precursors, simulants, and other common interfering substances on false positive alarms.
科研通智能强力驱动
Strongly Powered by AbleSci AI