Good Night: Sleep Monitoring Using a Physiological Radar Monitoring System Integrated with a Polysomnography System
作者
Mehran Baboli,Aditya Singh,Bruce Soll,Olga Borić-Lubecke,Victor M. Lubecke
出处
期刊:IEEE Microwave Magazine [Institute of Electrical and Electronics Engineers] 日期:2015-06-04卷期号:16 (6): 34-41被引量:64
标识
DOI:10.1109/mmm.2015.2419771
摘要
This article presents results from the integration of a noncontact physiological radar monitoring system (PRMS) with a type I polysomnography (PSG) system to perform sleep monitoring. The PRMS system consists of two continuous-wave Doppler radars operating at the industrial, scientific, and medical (ISM) band of 2.45 GHz. The system can acquire data, perform digital processing, and output appropriate conventional analog outputs with a latency of approximately 130 ms, which can be recorded and displayed by a gold standard sleep monitoring system along with other standard sensor measurements. Radar data was also used to successfully detect paradoxical motion that was simulated using linear movers and to categorize normal breathing, apnea, and hypopnea in sleeping subjects with obstructive sleep apnea (OSA).