OBJECTIVES: To evaluate the accuracy of CT-derived fat fraction (CDFF) software for quantifying hepatic steatosis at various radiation doses, using MRI-derived proton density fat fraction (MRI-PDFF) as the reference standard, and examines the impact of deep learning (DL)-based noise reduction on CDFF accuracy in low-dose CT (LDCT) scans. METHODS: We conducted a retrospective analysis of 125 living liver donor candidates who underwent non-contrast CT and MRI between July 2016 and December 2017. CDFF was measured on full-dose and simulated LDCT scans at 50%, 25%, and 10% radiation doses. DL-based denoising reconstruction (DLDR) was applied to LDCT scans for CDFF recalculation. The accuracy of CDFF was compared with MRI-PDFF using Pearson correlation coefficients and receiver operating characteristic (ROC) curve analysis, focusing on the effects of radiation dose and DLDR. RESULTS: Of the 125 participants (mean age 38 ± 10 years; 77 males), 29 (23%) had hepatic steatosis (MRI-PDFF ≥5%). Full-dose CDFF showed moderate correlation with MRI-PDFF (r = 0.728; P < .001). Correlation decreased with lower doses (r = 0.684-0.725) but improved with DLDR (r = 0.725-0.736). ROC AUC for diagnosing hepatic steatosis was 0.82 for full-dose CDFF, with similar performance across other doses except 10%. CONCLUSIONS: CDFF accuracy declines at lower radiation doses, but DLDR enhances accuracy, improving alignment with MRI-PDFF, especially at reduced doses. ADVANCES IN KNOWLEDGE: DLDR significantly enhances the accuracy of CDFF accuracy at lower radiation doses, enabling high diagnostic performance for hepatic steatosis while potentially reducing patient radiation exposure.
科研通智能强力驱动
Strongly Powered by AbleSci AI