Jae Yeon Hwang,Sung Woo Han,Seung Huh,Sohee Park,Sang Min Park,Jung Hoon Park
出处
期刊:Membranes
[Multidisciplinary Digital Publishing Institute]
日期:2025-08-26
卷期号:15 (9): 251-251
标识
DOI:10.3390/membranes15090251
摘要
The application of ceramic membranes in hemodialysis modules remains underexplored, as prior investigations have primarily concentrated on flat-sheet samples or small-scale assessments. This study advances the field by fabricating Al2O3 hollow fiber membranes, integrating them into a lab-scale module, and systematically evaluating the influence of sintering temperature on their structural characteristics, hemocompatibility, and dialysis performance. Al2O3 hollow fiber membranes were prepared using a phase inversion method and then sintered at three different temperatures. All membranes exhibited superior protein adsorption behavior compared to conventional polymer-based membranes, which indicates higher biocompatibility. Furthermore, the amount of adsorbed protein decreased with increasing sintering temperature. This suggests that the amount of protein adsorption can be controlled by adjusting the heat treatment conditions. The lab-scale hemodialyzer integrated with a membrane sintered at 1200 °C achieved the fastest urea removal rate of approximately 90% in 2 h and reached a Kt/V value of 1.1 after 60 min, which is comparable to the performance of commercial polymer-based hemodialyzers.