Erik Pettersson,Anna Bäck,Anne Thilander‐Klang
摘要
This study compares dual-energy computed tomography (DECT) images of a phantom including different material inserts and with additional lateral titanium or stainless steel inserts, simulating bilateral hip prostheses. Dual-source (DS) and fast kV-switching (FKS) DECT with/without metal artefact reduction (MAR) were compared with regards to virtually monoenergetic CT number accuracy and the depiction of different materials. Streak artefacts were observed between the metal inserts that were more severe with steel compared to titanium inserts. The artefact severity and CT number accuracy depended on the photon energy (keV) for both DECT techniques. While MAR generally increased the CT number accuracy and material depiction within the streak artefacts, it sometimes decreased the accuracy outside the streak artefacts for both DS and FKS. FKS depicted the metal inserts more accurately than DS with regards to both CT numbers and external diameter.