辅助
材料科学
支架
有限元法
弯曲
复合材料
结构工程
工程类
医学
外科
作者
Sukhwinder K. Bhullar,Hüseyin Lekesiz,Ahmet Abdullah Karaca,Yong Hyun Cho,Stephanie M. Willerth,Martin Byung‐Guk Jun
出处
期刊:Applied sciences
[Multidisciplinary Digital Publishing Institute]
日期:2022-01-17
卷期号:12 (2): 910-910
被引量:25
摘要
This study develops and characterizes the distinctive mechanical features of a stainless-steel metal stent with a tailored structure. A high-precision femtosecond laser was used to micromachine a stent with re-entrant hexagonal (auxetic) cell geometry. We then characterized its mechanical behavior under various mechanical loadings using in vitro experiments and through finite element analysis. The stent properties, such as the higher capability of the stent to bear upon bending, exceptional advantage at elevated levels of twisting angles, and proper buckling, all ensured a preserved opening to maintain the blood flow. The outcomes of this preliminary study present a potential design for a stent with improved physiologically relevant mechanical conditions such as longitudinal contraction, radial strength, and migration of the stent.
科研通智能强力驱动
Strongly Powered by AbleSci AI