有机磷
杀虫剂
生物传感器
生物
医学
毒理
生物化学
生态学
作者
Ani Mulyasuryani,Sasangka Prasetyawan
出处
期刊:Analytical chemistry insights
[SAGE]
日期:2015-01-01
卷期号:10: ACI.S30656-ACI.S30656
被引量:17
摘要
The research has developed an enzyme biosensor for the detection organophosphate pesticide residues. The biosensor consists of a pair of screen-printed carbon electrode (SPCEs). One of electrodes contains immobilized organophosphate hydrolase (OPH) on a chitosan membrane by cross-linking it with glutaraldehyde. The area of the electrodes was optimized to 3, 5, and 7 mm 2 . The OPH was isolated from Pseudomonas putida, and was purified by the ammonium sulfate precipitation method, with 6444 ppm (A) and 7865 ppm (B). The organophosphate pesticide samples were 0-100 ppb in tris-acetate buffer 0.05 M, pH 8.5. The results showed that the best performance of the biosensor was achieved by the enzyme A with an electrode area of 5 mm 2 . The sensitivity of the biosensor was between 3 and 32 µS/ppb, and the detection limit for the organophosphate pesticides was 40 ppb (diazinon), 30 ppb (malathion), 20 ppb (chlorpyrifos), and 40 ppm (profenofos).
科研通智能强力驱动
Strongly Powered by AbleSci AI