Milán Vecsey-Nagy,Giuseppe Tremamunno,Uwe Joseph Schoepf,Chiara Gnasso,Emese Zsarnóczay,Nicola Fink,Dmitrij Kravchenko,Moritz C. Halfmann,Jim O’Doherty,Bálint Szilveszter,Pál Maurovich‐Horvat,Ismail Mikdat Kabakus,Pal Spruill Suranyi,Tilman Emrich,Ákos Varga‐Szemes
出处
期刊:Radiology
[Radiological Society of North America]
日期:2025-03-01
卷期号:314 (3): e241479-e241479
被引量:44
标识
DOI:10.1148/radiol.241479
摘要
= .13). Total, calcified, and fibrotic plaque volumes demonstrated excellent agreement between repeated measurements and between readers for both PCD CT and EID CT (all intraclass correlation coefficients [ICCs] > 0.90). Whereas low-attenuation plaque volume had strong intrareader (ICC, 0.84; 95% CI: 0.57, 0.94) and interreader (ICC, 0.92; 95% CI: 0.81, 0.97) agreements for PCD CT, EID CT showed only moderate (ICC, 0.62; 95% CI: 0.11, 0.86) and poor (ICC, 0.47; 95% CI: 0.01, 0.79) intrareader and interreader reproducibility. Conclusion Compared with EID CT, PCD CT UHR imaging reduced segmented coronary plaque volume by nearly one-third and improved reproducibility of low-attenuation plaque measurements. © RSNA, 2025