己糖激酶
试剂
化学
色谱法
吸光度
葡萄糖-6-磷酸脱氢酶
脱氢酶
酶
固定化酶
葡萄糖氧化酶
流动注射分析
磷酸盐
生物化学
检出限
有机化学
糖酵解
作者
L P León,Douglas K. Chu,Lloyd Robert Snyder,Csaba Horváth
出处
期刊:Clinical Chemistry
[American Association for Clinical Chemistry]
日期:1980-01-01
卷期号:26 (1): 123-129
被引量:13
标识
DOI:10.1093/clinchem/26.1.123
摘要
Hexokinase and glucose-6-phosphate dehydrogenase, co-immobilized at the inner wall of plastic tubing, were used with continuous-flow analyzers to measure serum glucose. The performance of the enzyme-containing coils meets the hydraulic requirements of the continuous-flow instruments, and the average carryover does not exceed 2.2%. A glucose concentration change of 1 g/L yields a 0.08 change in absorbance. There is no significant deviation from linearity for glucose concentrations up to 5 g/L. The method is similar to the Reference Method proposed by the Food and Drug Administration, and results by the two methods agree and correlate well (r = 0.999). The operational stability of the enzyme tubes is sufficient to allow the analysis of at least 10 000 serum samples. We investigated in detail the effect of reagent concentrations and temperature on the performance of the enzyme coils, and the results shed light on the nature of the heterogeneous dual-enzymic reaction.