The MPT1 allows to determine the bubble dead time and to calculate the effective surface age of a bubble. The developed theory for describing adsorption processes at the surface of a growing bubble and useful approximate solutions of this theory give access to a quantitative interpretation of experiments.The standard version of the automated maximum bubble pressure tensiometer MPT1 has a time window from 2 ms up to 10 s. With a specially designed measuring cell of the MPT1 measurements down to 100 µs are possible.The capacity of the bubble pressure tensiometer is demonstrated by measurements of the dynamic surface tensions of aqueous solutions of different Tritons in the entire time interval. An analysis is given for the applicability of the bubble pressure method to extremely short adsorption times.