Matteo Forlini,Maå‚gorzata Szkaradek,Robert Ręjdak,Adriana Bratu,Paolo Maria Rossini,Domenico D’Eliseo,Cesare Forlini,Gian Maria Cavallini
出处
期刊:Retinal Cases & Brief Reports
[Ovid Technologies (Wolters Kluwer)]
日期:2017-01-01
卷期号:11 (3): 249-254
被引量:6
标识
DOI:10.1097/icb.0000000000000361
摘要
Purpose: To evaluate the efficacy of the modification of Adjustable Macular Buckling device in the treatment of myopic macular hole retinal detachment with posterior staphyloma. Methods: Four consecutive patients suffering from myopic macular detachment with macular hole were treated using the macular buckling procedure. An Adjustable Macular Buckling device was used in all four cases and was modified using a 29-gauge optical fiber to illuminate its macular plate. Results: Optical coherence tomography showed successful retinal reattachment and closure of the macular hole after the buckling procedure. The macular plate of the buckling device was properly positioned in all four patients. No complications were observed. Conclusion: The modification of the macular buckling device improves the accuracy of its positioning by illuminating its macular plate.