Ladan Yazdani,Manish Bhatt,Iman Rafati,An Tang,Guy Cloutier
出处
期刊:IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
[Institute of Electrical and Electronics Engineers]
日期:2022-04-11
卷期号:69 (6): 2061-2074
被引量:24
标识
DOI:10.1109/tuffc.2022.3166448
摘要
Ultrasound (US) shear wave (SW) elastography has been widely studied and implemented on clinical systems to assess the elasticity of living organs. Imaging of SW attenuation reflecting viscous properties of tissues has received less attention. A revisited frequency shift (R-FS) method is proposed to improve the robustness of SW attenuation imaging. Performances are compared with the FS method that we originally proposed and with the two-point frequency shift (2P-FS) and attenuation measuring US SW elastography (AMUSE) methods. In the proposed R-FS method, the shape parameter of the gamma distribution fitting SW spectra is assumed to vary with distance, in contrast to FS. Second, an adaptive random sample consensus (A-RANSAC) line fitting method is used to prevent outlier attenuation values in the presence of noise. Validation was made on ten simulated phantoms with two viscosities (0.5 and 2 Pa [Formula: see text]) and different noise levels (15 to -5 dB), two experimental homogeneous gel phantoms, and six in vivo liver acquisitions on awake ducks (including three normal and three fatty duck livers). According to the conducted experiments, R-FS revealed mean reductions in coefficients of variation (CV) of 62.6% on simulations, 62.5% with phantoms, and 62.3% in vivo compared with FS. Corresponding reductions compared with 2P-FS were 45.4%, 77.1%, and 62.0%, respectively. Reductions in normalized root-mean-square errors for simulations were 63.9% and 48.7% with respect to FS and 2P-FS, respectively.
科研通智能强力驱动
Strongly Powered by AbleSci AI