医学
动脉瘤
支架
数字减影血管造影
气球
放射科
血管成形术
颈内动脉
导管
血管造影
外科
作者
Keannette Russell,William C. Johnson,Fadi Al Saiegh,Lee Birnbaum,Alexander L. Coon,Justin Mascitelli
标识
DOI:10.1227/ons.0000000000001294
摘要
BACKGROUND AND IMPORTANCE: Giant aneurysms can present technical challenges during treatment with flow diversion including inability to access the aneurysm outflow directly. Encircling the aneurysm with a microwire/microcatheter has been well described; however, it can result in a twisted stent because of catheter twisting during the reduction maneuver, which, in turn, could lead to thromboembolic complications. CASE PRESENTATION: Here, we describe a novel technique to manage the twist of the flow diverter in a giant internal carotid artery aneurysm using a combination of angioplasty and off-label placement of a balloon-mounted cardiac stent within the flow diverter. CONCLUSION: At 1 year, the aneurysm is completely occluded on digital subtraction angiography and MRI, and the patient is neurologically intact.