D-二聚体
二聚体
抗体
免疫分析
聚乙二醇
医学
阻断抗体
化学
内科学
免疫学
生物化学
有机化学
作者
Hui Ma,Chunhua Chen,Yanmei Li,Jingjing Wang,Zhide Hu
标识
DOI:10.1016/j.cca.2022.08.011
摘要
Plasma D-dimer is a screening marker for thrombosis and hemostasis disorders. Falsely elevated D-dimer may result in misdiagnosis and potentially invasive investigations. Therefore, identifying falsely elevated D-dimer is of great value in laboratory practice. D-dimer is determined by immunoassay and vulnerable to heterophilic antibody interference. Here, we reported a case of falsely elevated D-dimer partially caused by heterophilic antibodies. Dilution test, polyethylene glycol precipitation, heterophilic blocking reagent and method comparison were used to identify the heterophilic antibody interference. All these methods, except for the heterophilic blocking reagent, revealed the existence of heterophilic antibody interference. Polyethylene glycol precipitation failed to reduce the D-dimer to its reference interval. Therefore, we conclude that the falsely elevated D-dimer is partially caused by heterophilic antibody interference.