医学
血流动力学
心肌缺血
心脏病学
缺血
内科学
作者
Sifan Chen,Qiao Li,Qinyu Pan,Qiuyan Yin,Liang Yue,Peng Zhang,G.-H. Chen,Weichao Liu
标识
DOI:10.1002/jbio.202300474
摘要
Noninvasive and real-time optical detection of cardiac hemodynamics dysfunction during myocardial ischemia remains challenging. In this study, we developed a near-infrared spectroscopy device to monitor rats' myocardial hemodynamics. The well-designed system can accurately reflect the hemodynamics changes by the classic upper limb ischemia test. Systemic hypoxia by disconnecting to the ventilator and cardiac ischemia by coronary artery slipknot ligation was conducted to monitor myocardial hemodynamics. When systemic hypoxia occurred, ΔHbR and ΔtHb increased significantly, whereas ΔHbO decreased rapidly. When coronary blood flow was obstructed by slipknots, cardiothoracic ΔHbO immediately begins to decline, while ΔHbR also significantly increases. Simultaneously, SpO
科研通智能强力驱动