色谱法
化学
检出限
二羟丙酮
高效液相色谱法
相关系数
乙腈
甘油
定量分析(化学)
有机化学
统计
数学
作者
Jie Chen,Jie Chen,Chen Zhou
标识
DOI:10.1093/chromsci/46.10.912
摘要
High-performance liquid chromatographic (HPLC) methods were respectively developed for the quantitative determination of dihydroxyacetone (DHA) and glycerol in the fermentation broth. Validation parameters such as linearity, precision, accuracy, and specificity, limit of detection (LOD), and limit of quantitation (LOQ) were determined. Both HPLC methods were carried out on a Lichrospher 5—NH2 column with a mobile phase constituted of acetonitrile and water (90:10, v/v). The linearity range for DHA was 2.00–12.00 mg/mL with a correlation coefficient (r) of 0.9994. The LOD and LOQ were 0.06 and 1.20 mg/mL, respectively. The linearity range for glycerol was 0.50–20.00 mg/mL with a correlation coefficient of 0.9998. The LOD and LOQ were 0.22 and 0.50 mg/mL, respectively. Also, the HPLC method to determine DHA was compared with an existing visible spectrophotometric method. Statistical analysis by F-test and t-test showed no significant difference at 95% confidence level between the two methods when applied to low DHA concentrations while a large deviation existed in the determinations of high DHA concentrations. The HPLC method was more accurate to determine high DHA concentrations.
科研通智能强力驱动