Yue Feng,Dan Z. Reinstein,Tore Nitter,Timothy J. Archer,Colm McAlinden,Geir Bertelsen,Aleksandar Stojanović
标识
DOI:10.3928/1081597x-20230606-01
摘要
Epithelial thickness mapping results were most repeatable with the Anterion, followed by the MS-39 and Avanti. The MS-39 gave the thickest values, followed by the Avanti and Anterion. The differences were significant, making the devices not interchangeable for epithelial thickness mapping in eyes with keratoconus. [J Refract Surg. 2023;39(7):474-480.].