The thickness of polyethylene tibial inserts has been implicated as a potential cause of excessive wear and early failure of total knee replacements. This paper highlights the fact that the information supplied by the manufacturers is inaccurate and potentially misleading; in one case, the true thickness was much less than the recommended minimum thickness. We recommend that the minimum thickness of the tibial components as well as the combined thickness of the polyethylene insert and the metal tibial tray be specified in the product information and on the packaged insert.