角膜塑形术
延伸率
医学
眼科
验光服务
材料科学
角膜
极限抗拉强度
冶金
作者
Mark A. Bullimore,Maria Liu
标识
DOI:10.1016/j.clae.2023.101875
摘要
The Euclid Emerald lens designs for orthokeratology have been available in global markets for over 20 years and is used extensively by clinicians for slowing myopia progression in children. This paper comprehensively reviews data from published studies of the efficacy of this lens.A comprehensive systematic search was performed in March 2023 using Medline with the following search terms: orthokeratology AND myopi* AND (axial or elong*) NOT (review or meta).The original search identified 189 articles, of which 140 reported axial elongation. Of those, 49 reported data on the Euclid Emerald design. Unique axial elongation data could be extracted from 37 papers-14 of which included an untreated control group. Among these, the mean 12-month efficacy-the difference in axial elongation between orthokeratology wearers and controls-was 0.18 mm (range: 0.05-0.29 mm), and the mean 24-month efficacy was 0.28 mm (range: 0.17-0.38 mm). The orthokeratology wearers in 23 studies without an untreated comparison group showed similar axial elongation to those in the 14 studies with a control group. For example, the mean 12-month axial elongation for the studies with controls was 0.20 ± 0.06 mm compared with 0.20 ± 0.07 mm for the studies without controls.This extensive body of literature on a single device for myopia control is unique and demonstrates the efficacy of this design in slowing axial elongation in myopic children.
科研通智能强力驱动