吸气器
医学
镊子
失血
外科
肝切除术
切除术
热力学
物理
作者
Nobuhiko Taniai,Masahiko Onda,Takashi Tajiri,Koho Akimaru,Hiroshi Yoshida,Yasuhiro Mamada
出处
期刊:PubMed
日期:2002-10-26
卷期号:49 (48): 1649-51
被引量:7
摘要
We introduced an ultrasonic surgical aspirator with electrosurgical coagulation to increase safety of the liver resections performed for hepatocellular carcinoma.This system was evaluated by analyzing the intraoperative blood loss, the duration of hepatectomy, the ischemia time, and the postoperative clinical course and comparing these same parameters with the traditional forceps' fracture method.There was no significant difference in the duration of surgery and the mean liver resection time between the two methods. The mean blood loss by using this system was 2458 +/- 4742 mL and by forceps' fracture method 956 +/- 252 mL. The mean ischemia time was 44.4 +/- 35.9 min by using this system and 23.9 +/- 29.1 min in forceps' fracture method. There was a significant decrease in intraoperative blood loss and ischemia time using the new system. This may decrease postoperative complications.This system may enable all surgeons to perform liver resection easily and safely.