医学
观察研究
怀孕
糖尿病
闭环
多中心研究
2型糖尿病
妊娠期糖尿病
产科
1型糖尿病
循环(图论)
妊娠期
内科学
内分泌学
随机对照试验
妊娠期糖尿病
控制工程
工程类
组合数学
生物
遗传学
数学
作者
Carol J. Levy,Yogish C. Kudva,Başak Özaslan,Kristin Castorino,Grenye O’Malley,Ravinder Jeet Kaur,Camilla Levister,Mei Mei Church,Donna Desjardins,Shelly McCrady-Spitzer,Selassie Ogyaadu,Mari Charisse Trinidad,Corey Reid,Shafaq Rizvi,Sunil Deshpande,Isabella Zaniletti,Walter K. Kremers,Jordan E. Pinsker,Francis J. Doyle,Eyal Dassau
出处
期刊:Diabetes Care
[American Diabetes Association]
日期:2023-05-17
卷期号:46 (7): 1425-1431
被引量:32
摘要
OBJECTIVE There are no commercially available hybrid closed-loop insulin delivery systems customized to achieve pregnancy-specific glucose targets in the U.S. This study aimed to evaluate the feasibility and performance of at-home use of a zone model predictive controller–based closed-loop insulin delivery system customized for pregnancies complicated by type 1 diabetes (CLC-P). RESEARCH DESIGN AND METHODS Pregnant women with type 1 diabetes using insulin pumps were enrolled in the second or early third trimester. After study sensor wear collecting run-in data on personal pump therapy and 2 days of supervised training, participants used CLC-P targeting 80–110 mg/dL during the day and 80–100 mg/dL overnight running on an unlocked smartphone at home. Meals and activities were unrestricted throughout the trial. The primary outcome was the continuous glucose monitoring percentage of time in the target range 63–140 mg/dL versus run-in. RESULTS Ten participants (HbA1c 5.8 ± 0.6%) used the system from mean gestational age of 23.7 ± 3.5 weeks. Mean percentage time in range increased 14.1 percentage points, equivalent to 3.4 h per day, compared with run-in (run-in 64.5 ± 16.3% versus CLC-P 78.6 ± 9.2%; P = 0.002). During CLC-P use, there was significant decrease in both time over 140 mg/dL (P = 0.033) and the hypoglycemic ranges of less than 63 mg/dL and 54 mg/dL (P = 0.037 for both). Nine participants exceeded consensus goals of above 70% time in range during CLC-P use. CONCLUSIONS The results show that the extended use of CLC-P at home until delivery is feasible. Larger, randomized studies are needed to further evaluate system efficacy and pregnancy outcomes.
科研通智能强力驱动
Strongly Powered by AbleSci AI