杀伤力
过氧化氢
过程(计算)
气相
相(物质)
化学
材料科学
环境科学
计算机科学
生物
毒理
生物化学
有机化学
热力学
操作系统
物理
标识
DOI:10.5731/pdajpst.2024-003033.1
摘要
This manuscript reviews the similarities between death curves associated with 'rogue' biological indicator results in vapor phase hydrogen peroxide decontamination / sterilization processes and those expected from varying process conditions. The manuscript considers whether the perceived 'rogue' behavior of some biological indicators in these process may be the result of the phase differences. Consideration of 'wet' vs 'dry' processes are discussed and recommendations for process improvement to reduce failures are provided.
科研通智能强力驱动
Strongly Powered by AbleSci AI