氟磷灰石
羟基磷灰石
材料科学
植入
钛
扫描电子显微镜
合金
钛合金
生物材料
磷灰石
复合材料
生物医学工程
冶金
化学
外科
医学
矿物学
生物化学
纳米技术
酶
作者
Wouter J.A. Dhert,Christel P. A. T. Klein,Joop G.C. Wolke,Edo A. van der Velde,K. de Groot,P. M. Rozing
出处
期刊:Journal of Biomedical Materials Research
[Wiley]
日期:1991-10-01
卷期号:25 (10): 1183-1200
被引量:167
标识
DOI:10.1002/jbm.820251002
摘要
Ceramic coatings of fluorapatite (FA), magnesiumwhitlockite (MW), and hydroxylapatite (HA), and noncoated Ti-6Al-4V alloy (Ti) implants were evaluated before and after implantation in an animal study. Cylindrical plugs were coated by plasma-spraying with FA, MW, and HA. X-ray-diffraction patterns showed for FA and HA a crystalline and for MW an amorphous-crystalline coating structure. The plugs were implanted into the right femora and left humeri of 16 adult goats. Follow-up periods were 12 and 25 weeks. The in vivo results were evaluated using push-out tests and scanning electron microscopy. There were significant differences in push-out strengths between femur and humerus. The FA and HA implants showed significantly higher push-out strengths than the MW and Ti alloy implants, especially for the 12 week follow-up period. Furthermore, at 12 weeks, MW showed significantly lower push-out strengths than Ti alloy. SEM-investigation of the interfaces revealed that FA did not degrade while both MW and HA showed extensive degradation at 12 and 25 weeks. In addition, the interface after push-out testing showed for the MW, HA, and Ti alloy implants to be fractured at the implant-tissue interface and for the FA to be fractured at the coating-titanium interface.
科研通智能强力驱动