接口
无线
计算机科学
脑植入物
计算机硬件
频道(广播)
植入
脑-机接口
嵌入式系统
电气工程
工程类
电信
医学
人工智能
外科
脑电图
精神科
作者
C. Mestais,Guillaume Charvet,Fabien Sauter-Starace,Michael Foerster,David Ratel,Alim Louis Benabid
标识
DOI:10.1109/tnsre.2014.2333541
摘要
A wireless 64-channel ElectroCorticoGram (ECoG) recording implant named WIMAGINE has been designed for various clinical applications. The device is aimed at interfacing a cortical electrode array to an external computer for neural recording and control applications. This active implantable medical device is able to record neural activity on 64 electrodes with selectable gain and sampling frequency, with less than 1 μV(RMS) input referred noise in the [0.5 Hz - 300 Hz] band. It is powered remotely through an inductive link at 13.56 MHz which provides up to 100 mW. The digitized data is transmitted wirelessly to a custom designed base station connected to a PC. The hermetic housing and the antennae have been designed and optimized to ease the surgery. The design of this implant takes into account all the requirements of a clinical trial, in particular safety, reliability, and compliance with the regulations applicable to class III AIMD. The main features of this WIMAGINE implantable device and its architecture are presented, as well as its functional performances and long-term biocompatibility results.