作者
Qiwen Sheng,Xiaoli Wang,Musaddeque Anwar Al-Abedin Syed,Hasanur Rahman Chowdhury,Seungsup Lee,Andrew Dvorak,M.W. Shafer,Ming Han
摘要
Fiber-optic bolometers (FOBs) designed for plasma radiation measurement in magnetically confined fusion environments have been previously developed and tested at the DIII-D tokamak. These FOBs utilize a silicon Fabry–Perot interferometer at the fiber tip for temperature measurement and a gold disk as a radiation absorber. This paper presents recent advancements in the development of FOBs with reduced cooling time constants and enhanced temperature tolerance. We also demonstrate high-temperature operation and report findings from irradiation tests. By incorporating a copper sensor holder as a thermal sink and integrating two closely packed sensor heads into a compact design, the cooling time constants were reduced to ∼100 ms in vacuum, while minimizing uncertainties caused by temperature gradients. The FOBs were subjected to heating at 400 °C for five hours, with no observed degradation in noise performance post-heating. High-temperature operation tests were conducted at ∼220 °C using a 404 nm square-wave modulated laser to simulate the plasma radiation. In addition, four FOBs constructed from copper-coated pure-silica fibers were fabricated and underwent irradiation testing, enduring a total γ-radiation dose of 15.2 MGy and a neutron fluence of 1.6 × 1018 n/cm2. The FOBs maintained their physical integrity and the optical signal shows high visibility throughout the test.