闪烁体
中子成像
中子探测
中子
物理
光学
锌
探测器
核物理学
材料科学
冶金
作者
Bin Tang,Yang Chang Wu,Hang Li,Yong Sun,Heyong Huo,Bin Liu,Ke Tang,Wei Hsian Yin,Chao Chao
标识
DOI:10.1016/j.nima.2013.07.063
摘要
Abstract Fast neutron radiography is a promising application for accelerator based neutron sources. The potential effectiveness of this technique depends on the development of suitable imaging detectors for fast neutrons. Zinc sulphide based scintillators have the largest light output per event in the family of imaging scintillators used so far in fast neutron radiography. This paper investigated different aspects of this scintillator in order to determine the factors which might affect the light output. A mathematical model was established to estimate effectiveness of this scintillator. Zinc sulphide screens were prepared with ZnS particles of different concentrations in polypropylene matrix. A 14 MeV fast neutron source was used in the experiments. The light output was detected using a CCD camera or a film coupled to the scintillator screen. The results showed that the optimum scintillators is around 3 mm in thickness with the weight ratio of 2:1 for ZnS and polypropylene.
科研通智能强力驱动
Strongly Powered by AbleSci AI