作者
Stephen Zambrzycki,Vilmos Kertész,John F. Cahill
标识
DOI:10.1021/jasms.5c00405
摘要
High-throughput single-cell mass spectrometry is a rapidly evolving field that requires innovative sampling and ionization techniques to balance speed, sensitivity, and reliability for metabolomic and lipidomic analyses. This study provides a comparative analysis of two cutting-edge ionization platforms for single-cell analysis: Liquid Vortex Capture (LVC) and Rapid Droplet Sampling Interface (RDSI). The performance was benchmarked by testing pharmaceuticals, EquiSPLASH, and single-cell experiments. RDSI demonstrated up to 100-fold improvements in sensitivity for drugs and lipids such as propranolol, amiodarone, atorvastatin, and phosphocholines in water and phosphate-buffered solutions. This was attributed to its low-flow rate operation (3 μL/min) and reduced dilution. Conversely, LVC excelled in handling higher liquid volumes with greater reproducibility due to its higher solvent flow rate (200 μL/min), enabling increased dilution, solubility, and cleaning. Single-cell uptake of atorvastatin incubated for 10 min, or amiodarone incubated for 24 h in HepG2 cells, similarly revealed up to 85-fold enhancement in sensitivity by RDSI for drugs and lipids. These findings highlight the potential of RDSI for enhancing sensitivity in single-cell drug monitoring and lipidomics.