Scanning force microscopy with micromachined silicon sensors
作者
Martin Nonnenmacher,Johann Greschner,Olaf Wolter,Rainer Kassing
出处
期刊:Journal of vacuum science & technology [American Institute of Physics] 日期:1991-03-01卷期号:9 (2): 1358-1362被引量:112
标识
DOI:10.1116/1.585196
摘要
Scanning force microscopy (SFM), operated in the attractive imaging mode, enables the precise measurement of the force between tip and sample over a tip–sample distance ranging from contact to tens of nanometers. The basic long range interactions (>1 nm: i.e., hydrodynamic, electrostatic, van der Waals, and capillary forces) between tip and sample have been measured and will be discussed. Each force leads to a different mode of operation in profiling samples. The most critical part of the SFM is the force sensor. Exact knowledge of the sensor properties is required for the interpretation of SFM measurements. We have used micromachined silicon sensors consisting of a monolithic silicon cantilever with integrated silicon tip and have performed a detailed characterization of the tip geometry and resonance properties. Examples of surface images on different samples (conductors, insulators and biological materials) and structures, ranging from atomic steps up to several microns high features, have been investigated to demonstrate capabilities and problems in SFM imaging.