化学
离子阱
质谱法
串联质谱法
串联
四极离子阱
碰撞诱导离解
分析化学(期刊)
离子
选择性反应监测
离解(化学)
电离
电子转移离解
分光计
色谱法
光学
航空航天工程
工程类
物理
有机化学
物理化学
作者
Xinwei Liu,Xiao Wang,Jiexun Bu,X. K. Zhou,Zheng Ouyang
标识
DOI:10.1021/acs.analchem.8b03958
摘要
Miniature mass spectrometers are of an increasing interest for in-situ analyses, and their coupling with the ambient ionization sources is a valid path for direct analysis of complex samples. In this study, a miniature mass spectrometer using discontinuous atmospheric pressure interface was developed with a dual-LIT (linear ion trap) configuration. The comprehensive scan modes were enabled for tandem mass spectrometry analysis, which are critical for high-quality qualitative and quantitative analysis. A real-time pressure control was implemented to facilitate the ion transfer and collision induced dissociation (CID). Beam-type CID could be performed for tandem analysis at a high number of stages. In-trap CID at high q could also be performed with the fragment ions accumulated in a second trap. A precursor ion scan mode for analyzing target analytes has also been demonstrated.
科研通智能强力驱动
Strongly Powered by AbleSci AI