热成像
电容描记术
红外线的
呼吸
金标准(测试)
生物医学工程
医学
麻醉
内科学
物理
光学
作者
Carina Barbosa Pereira,Xinchi Yu,Vladimír Blažek,Steffen Leonhardt
标识
DOI:10.1109/embc.2015.7319333
摘要
An abnormal breathing rate (BR) is one of the strongest markers of physiological distress. Moreover, it plays an important role in early detection of sudden infant death syndrome, as well as in the diagnosis of respiratory disorders. However, the current measuring modalities can cause discomfort to the patient, since attachment to the patient's body is required. This paper proposes a new approach based on infrared thermography to remotely monitor BR. This method allows to (1) detect automatically the nose, (2) track the associate region of interest (ROI), and (3) extract BR. To evaluate the performance of this method, thermal recording of 5 healthy subjects were acquired. Results were compared with BR obtained by capnography. The introduced approach demonstrated an excellent performance. ROIs were precisely segmented and tracked. Furthermore, a Bland-Altman diagram showed a good agreement between estimated BR and gold standard. The mean correlation and mean absolute BR error are 0.92 ± 0.07 and 0.53 bpm, respectively. In summary, infrared thermography seems to be a great, clinically relevant alternative to attached sensors, due to its outstanding characteristics and performance.