This work presents a planar transformer-based balun for magnetic resonance imaging (MRI) transmit systems at 128 MHz and 3 kW. The design uses single-turn primary and secondary coils on separate printed circuit board layers, with magnetic coupling through dielectric and air, eliminating ferromagnetic materials. A 4 : 1 impedance transformation matches a differential transistor output impedance of 6.25 Ω per drain to 50 Ω, while converting to a single-ended output. Measurements show 14.8 dB return loss, 0.3 dB insertion loss, amplitude/phase imbalance below 0.1 dB/5°, and 27.5 dB common-mode rejection. Thermal results confirm stable operation. The balun provides an efficient solution for impedance matching and differential-to-single-ended conversion in MRI transmit amplifiers.