扫描电化学显微镜
化学
扫描离子电导显微镜
电极
显微镜
循环伏安法
微电极
纳米环
超微电极
分析化学(期刊)
电化学
纳米技术
材料科学
光学
扫描共焦电子显微镜
物理
物理化学
色谱法
作者
Yasufumi Takahashi,Andrew Shevchuk,Pavel Novák,Yumi Murakami,Hitoshi Shiku,Yuri E. Korchev,Tomokazu Matsue
摘要
We described a hybrid system of scanning electrochemical microscopy (SECM) and scanning ion conductance microscopy (SICM) with ion current feedback nanopositioning control for simultaneous imaging of noncontact topography and spatial distribution of electrochemical species. A nanopipette/nanoring electrode probe provided submicrometer resolution of the electrochemical measurement on surfaces with complex topology. The SECM/SICM probe had an aperture radius of 220 nm. The inner and outer radii of the SECM Au nanoring electrode were 330 and 550 nm, respectively. Characterization of the probe was performed with scanning electron microscopy (SEM), cyclic voltammetry (CV), and approach curve measurements. SECM/SICM was applied to simultaneous imaging of topography and electrochemical responses of enzymes (horse radish peroxidase (HRP) and glucose oxidase (GOD)) and single live cells (A6 cells, superior cervical ganglion (SCG) cells, and cardiac myocytes). The measurements revealed the distribution of activity of the enzyme spots on uneven surfaces with submicrometer resolution. SECM/SICM acquired high resolution topographic images of cells together with the map of electrochemical signals. This combined technique was also applied to the evaluation of the permeation property of electroactive species through cellular membranes.
科研通智能强力驱动
Strongly Powered by AbleSci AI