校准
质谱法
环境科学
遥感
化学
色谱法
地质学
数学
统计
作者
Cong Mei,Chen Ming Fu,Wei Yan Zhang,Jiangkun Li,Xudong Liang,Wu Xue,Qing-Fa Luo,Yaxin Yang
标识
DOI:10.1016/j.net.2025.103594
摘要
The calibration facility process for the airborne γ-spectrometer, which is built at the Shijiazhuang Dagou Village Airport of China, is time-consuming and labor-intensive. In this paper, a sourceless efficiency calibration model based on the Monte Carlo method using the Geant4 software has been established. The proposed model allows for the acquisition of the γ-spectrum response curve of the airborne γ-spectrometer above a cylindrical γ-radiation source containing uranium, thorium, and potassium elements in any proportion. It is applied to perform Compton scattering calibration and height attenuation correction for UGRS-10 airborne gamma spectrometer. The relative errors between the calibration parameters obtained by Geant4 simulation and those obtained by UGRS-10 airborne gamma spectrometer at Daguocun airport calibration facility are between 3 % and 27 %. The experiment demonstrates the Monte Carlo-based sourceless efficiency calibration model is suitable to calibrate airborne γ-spectrometers and provides technical support for airborne γ-spectrometry.
科研通智能强力驱动
Strongly Powered by AbleSci AI