The gold electrode was modified with cupric mercapto-butadiamine(abbreviated as CuL) and used as a chemical sensor in the immuno-assay of the antigen HIV-p24.The sensor was self-assembled with a monolayer of CuL,giving a surface coverage of 1.02×10-11mol·cm-2.Calibration curve was prepared and a linearity in the range of 0.5 ng to 150 μg per L of HIV-p24 against the voltammetric response,in a phosphate buffer solution of pH 6.2,and containing 1.0 mmol·L-1 of H2O2 was obtained.The detection limit found was 0.2 μg·L-1.Precision was tested with a same electrode at the concentration level of 10 μg·L-1 of HIV-p24,and the RSD obtained was 3.5%.Recovery was tested by addition of HIV-p24 at known concentrations of 5,15,20 and 30 μg·L-1,and the values of recovery found were in the range of 98%-102%.The addition of electron transfer mediator to the test solution was not necessary in the proposed method.In the application of this method to clinical analysis of serum samples,the results obtained were in consistency with those obtained by the radio-immunoassay,and the RSD′s(n=7) obtained were less than 0.3%.