BackgroundAs one of the most important issues in the field of gamma spectrum analysis, the main function of spectrum stabilization is to suppress the "spectral line drift" caused by environmental changes. Portable LaBr3(Ce) gamma spectrometer is mainly used to analyze and process the gamma spectra of radionuclides and obtain qualitative and quantitative information about radionuclides. It often works in the field complex environment where the spectrum drift is inevitable, hence must have the function of spectrum stabilization.PurposeThis study aims to improve the performance of spectrum stabilization technique for a portable LaBr3(Ce) gamma spectrometer.MethodsThe intrinsic characteristic peak of LaBr3(Ce) detector and the peak of 40K in the natural background were made of as the reference peaks. Then, the concept of "spectrum similarity" was defined to calculate the similarity between the measured spectrum and the standard background spectrum. Finally, a new spectrum stabilization method using standard reference peak and "spectrum similarity" was proposed to measure and correct the deviation. Experiments were carried out verify the performance of this new method.ResultsExperimental results show that the technology can effectively stabilize spectrum in the temperature range of -30 ℃ to 50 ℃ without calibration of the relationship curve between the reference peak and the temperature, and the peak drift of the reference peak is stable within ±1 channel even if the temperature changes more than 50 ℃∙h-1.ConclusionsThis technology effectively improves the application range of portable LaBr3(Ce) spectrometers.