摘要
The principal component of any non-invasive blood pressure measurement system is an inflatable cuff. Different types of fabrics are used for inflatable cuffs construction. In this study, sphygmomanometric blood pressure measurement using inflatable cuffs was simulated in Abaqus and validated through experimental results. The purpose of the simulation is to study the effect of variation in cuff fabric geometric and mechanical properties on pressure distribution and pressure transmission during blood pressure measurement by predicting the pressure at the interface of the blood pressure cuffs and a metal cylinder. Geometric and mechanical properties of the fabrics of four different cuff types were found experimentally. Interface pressure at the cuffs and metal cylinder surface was also found experimentally using Tekscan pressure sensing system for models validation. The results of the simulation showed that the interface pressure underneath the cuffs vary with variation in geometric and mechanical properties of their fabrics. The results of the simulation were found to be in good agreement with experimental findings. This research demonstrates that the pressure distribution under the cuffs is related to the cuffs' fabric geometric and mechanical properties. This means that variation in cuffs' fabric properties could ultimately incur variations in the blood pressure values of human subjects.