三氟乙酸
电极
色谱法
化学
检出限
校准曲线
校准
高效液相色谱法
数学
统计
物理化学
作者
Santiago Cavanillas,Núria Serrano,José Manuel Díaz‐Cruz,Cristina Ariño,Miquel Esteban
标识
DOI:10.1002/elan.201300593
摘要
Abstract In order to show the possibilities of screen‐printed electrodes in HPLC detection, a method is presented for routine quantification of aminothiols in human plasma. It uses commercial gold screen‐printed electrodes and provides a rapid, simple and economical procedure with a disposable detector and without any additional electrode. By working at 98 : 2 (0.05 % trifluoroacetic acid: methanol) and 0.9 V, it allows us to determine cysteine (Cys), cysteinyl‐glycine (Cys‐Gly), and homocysteine (HCys) at μmol/L levels. The method was validated, and linearity range, detection and quantification limits, precision and accuracy were evaluated by external calibration and by using N ‐acetylcysteine as internal standard, which exhibited better precision.