多导睡眠图
就寝时间
睡眠(系统调用)
物理疗法
医学
听力学
心理学
物理医学与康复
呼吸暂停
麻醉
内科学
计算机科学
操作系统
作者
Andrée‐Anne Parent,Veronica Guadagni,Jean M. Rawling,Marc J. Poulin
出处
期刊:Sensors
[Multidisciplinary Digital Publishing Institute]
日期:2024-03-30
卷期号:24 (7): 2218-2218
被引量:4
摘要
Introduction: This study aimed to validate the ability of a prototype sport watch (Polar Electro Oy, FI) to recognize wake and sleep states in two trials with and without an interval training session (IT) 6 h prior to bedtime. Methods: Thirty-six participants completed this study. Participants performed a maximal aerobic test and three polysomnography (PSG) assessments. The first night served as a device familiarization night and to screen for sleep apnea. The second and third in-home PSG assessments were counterbalanced with/without IT. Accuracy and agreement in detecting sleep stages were calculated between PSG and the prototype. Results: Accuracy for the different sleep stages (REM, N1 and N2, N3, and awake) as a true positive for the nights without exercise was 84 ± 5%, 64 ± 6%, 81 ± 6%, and 91 ± 6%, respectively, and for the nights with exercise was 83 ± 7%, 63 ± 8%, 80 ± 7%, and 92 ± 6%, respectively. The agreement for the sleep night without exercise was 60.1 ± 8.1%, k = 0.39 ± 0.1, and with exercise was 59.2 ± 9.8%, k = 0.36 ± 0.1. No significant differences were observed between nights or between the sexes. Conclusion: The prototype showed better or similar accuracy and agreement to wrist-worn consumer products on the market for the detection of sleep stages with healthy adults. However, further investigations will need to be conducted with other populations.
科研通智能强力驱动
Strongly Powered by AbleSci AI