化学
检出限
呼气
热脱附
质谱法
分析化学(期刊)
色谱法
气体分析呼吸
解吸
体积流量
粒径
挥发性有机化合物
粒子(生态学)
微量气体
采样(信号处理)
作者
Jie Jin,Wei Xu,Ning Zhang,Qi Zhang,Xun Bao,Qu Liang,Qiangling Zhang,Chaoqun Huang,Xue Zou,Chengyin Shen,Yannan Chu
出处
期刊:Analyst
[Royal Society of Chemistry]
日期:2026-04-20
卷期号:151 (8): 2290-2298
摘要
headspace sampling into a proton transfer reaction-mass spectrometer (PTR-MS) for detection. This study optimized parameters of the LTE-HS-PTR-MS system, including the type, total volume, particle size of the enrichment material, and pre-concentration flow rate. The system's performance in terms of enrichment efficiency, sensitivity, limit of detection (LOD), and recovery rate was evaluated. The results showed that this technique could reduce the instrument's LOD by 23.4 to 28.8 times. Finally, comparative analysis of breath samples from healthy individuals using two detection techniques revealed that, compared to conventional PTR-MS, LTE-HS-PTR-MS significantly enhanced the signal intensity of 17 product ions from exhaled VOCs, with an average enrichment efficiency exceeding 29 times. The highly sensitive detection technique developed in this study facilitates the detection of trace VOCs in exhalation and holds potential for the discovery of breath biomarkers for diseases.
科研通智能强力驱动
Strongly Powered by AbleSci AI