热电偶
材料科学
电磁屏蔽
电阻式触摸屏
涡流
屏蔽电缆
温度测量
电流(流体)
声学
电气工程
核磁共振
机械
物理
复合材料
工程类
量子力学
作者
Dev P. Chakraborty,I Brezovich
出处
期刊:The Journal of microwave power
[Informa]
日期:1982-01-01
卷期号:17 (1): 17-28
被引量:48
标识
DOI:10.1080/16070658.1982.11689261
摘要
Thermocouple thermometry errors in radiofrequency (typically 13, 56 MHZ) electromagnetic fields such as are encountered in hyperthermia are described. RF currents capacitatively or inductively coupled into the thermocouple-detector circuit produce errors which are a combination of interference, i.e., 'pick-up' error, and genuine rf induced temperature changes at the junction of the thermocouple. The former can be eliminated by adequate filtering and shielding; the latter is due to (a) junction current heating in which the generally unequal resistances of the thermocouple wires cause a net current flow from the higher to the lower resistance wire across the junction, (b) heating in the surrounding resistive material (tissue in hyperthermia), and (c) eddy current heating of the thermocouple wires in the oscillating magnetic field. Low frequency theories are used to estimate these errors under given operating conditions and relevant experiments demonstrating these effects and precautions necessary to minimize the errors are described. It is shown that at 13.56 MHz and voltage levels below 100 V rms these errors do not exceed 0.1 degrees C if the precautions are observed and thermocouples with adequate insulation (e.g., Bailey IT-18) are used. Results of this study are being currently used in our clinical work with good success.