Measurement uncertainty analysis of the electro-optical E-field measurement system
作者
Xu Kong,Yan‐Zhao Xie
标识
DOI:10.1109/ceem.2015.7368689
摘要
The measurement uncertainty of the electro-optical E-field measurement system is analyzed and calculated in this paper. Firstly, by analyzing the measurement and calibration process, the mathematical model of the measurement system is proposed. Then the influence factors of the measuring result is listed and their individual uncertainties are calculated under some approximations and assumptions. Finally, the estimate of the relative expanded measurement uncertainty is 20%, under the confident level of 95%.