Delaram Jarchi,Dario Salvi,Carmelo Velardo,Adam Mahdi,Lionel Tarassenko,David A. Clifton
标识
DOI:10.1109/bsn.2018.8329679
摘要
We describe an evaluation of photoplethysmography (PPG) signals with two wavelengths channels (infrared and red) using a wrist-worn sensor for the estimation of heart rate variability (HRV) and oxygen saturation (SpO2). Five healthy subjects were equipped with a commercial wrist-worn pulse oximeter (Wavelet Health, USA) on the right hand, and both a commercial smart watch for fitness use (Huawei Watch, Series 2) and a clinically-validated transmission-mode pulse oximeter (Creative Medical PC-68B) on their left hand as a reference. Synchronised PPG signals from the Wavelet Health, the Huawei watch, and the PC-68B were recorded for approximately 10 minutes. Subjects were asked to leave the left hand in a resting state, while moving the right hand with two types of movement (periodic and random). A method is proposed to incorporate coupling information between the two wavelengths of PPGs based on the bivariate empirical mode decomposition algorithm. Our method is shown to improve the quality of red PPG allowing improved estimation of SpO2. A comparison of average heart rate (HR), HRV, and SpO2 from all devices is provided.