Clive R. Hamlin,Dan A. Pankowsky
出处
期刊:Clinical Chemistry
[American Association for Clinical Chemistry]
日期:1987-01-01
卷期号:33 (1): 144-146
被引量:14
标识
DOI:10.1093/clinchem/33.1.144
摘要
We describe a simple, rapid turbidimetric method for measuring transthyretin in human serum by use of a centrifugal analyzer (Cobas-Bio) and commercially available antiserum. The immunoturbidimetric reaction is monitored at 340 nm for 180 s, and the changes in absorbance after the antiserum is added are used to prepare a stable standard curve. The range of measurement is 10 to 520 mg/L. The CV for between-day precision was less than 5.0% for normal and low controls. Bilirubin up to 310 mumol/L and hemoglobin up to 225 mumol/L did not interfere, but triglycerides greater than 2.3 mmol/L falsely depress the results. The normal reference interval for adults, derived from data on subjects 20 to 65 years old (both sexes), is 193 to 376 mg/L. Concentrations are lower in children (67-302 mg/L) and in adults 65 or more years old (113-292 mg/L).