Na Li,Zhongchen Wu,Yingying Wang,Jing Zhang,Xiangnan Zhang,Hengnan Zhang,Wenhai Wu,Jing Gao,Jie Jiang
标识
DOI:10.1021/acs.analchem.6b03523
摘要
This paper describes the first demonstration of a portable dielectric barrier discharge-atomic emission spectrometer (DBD-AES). The instrument primarily consists of a miniature electro-thermal vaporizer (ETV), DBD, and optical signal acquisition units. It weighs only 4.5 kg and is powered by a 24 V DC battery with a maximum power consumption of 37 W. The accompanying software can be operated on a laptop computer. A specially designed quartz tube integrates the ETV unit with the DBD chamber. The effects of experimental parameters were investigated. The limit of detection (LOD) for mercury was 0.4 μg L–1 (1.2 pg) with a sampling volume of 3 μL. The instrument is applicable for multielement analysis, and the LODs ranged from 0.16 to 11.65 μg L–1 for Zn, Pb, Ag, Cd, Au, Cu, Mn, Fe, Cr, and As. The instrument was also validated by in-field analysis of seawater samples. The experimental results demonstrated the sensitivity, reliability, and practicality of the instrument.
科研通智能强力驱动
Strongly Powered by AbleSci AI