A sensor based on mercapto-butadiamine copper(CuL) monolayer modified gold electrode for determination of HIV-p24 antigen in human serum is proposed.The main parameters of the sensor are as follows.Diameter of the electrode surface is 2 mm and the height is 2 cm.CuL is self-assembled on Au electrode which gives a surface coverage of 1.02×10-11 mol/cm2.The electrode can detect HIV-p24 antigen within 10 s for blood volume less than 10 mL.The electrochemistry of the immobilized CuL displays a surface-controlled electrode process.Based on an electrochemical enzyme-linked immunoassay,the calibration curve was obtained ranging from 0.5 to 200 ng/mL and the detection limit is 0.2 ng/mL for HIV-p24 determination in pH=6.2 phosphate buffer solution containing 8 mmol/L H2O2.The sensor shows good accuracy,precision and reproducibility with the coefficient of variation of 3.5% at 10 ng/mL HIV-p24 concentration.This method avoids the addition of electron transfer mediator to the solution.The proposed method would be valuable for the diagnosis of AIDS disease.