氙气
氦
电离
氦离子探测器
闪烁
放电电离检测器
产量(工程)
电离室
原子物理学
激发态
化学
分析化学(期刊)
材料科学
探测器
物理
离子
光学
有机化学
冶金
色谱法
作者
Akihiro Takeuchi,Kiwamu Saito,Y. Kishimoto,Takahiro Oyama,Toshiya Sanami
标识
DOI:10.35848/1347-4065/ab7a47
摘要
The scintillation and ionization yields of helium–xenon gas mixture were measured using a gridded ionization chamber with 241 Am alpha particles to study the potential of the mixture to be applied in neutron detectors. The ionization yields for various ratios of the mixture gases were obtained with gas pressures of 1.0, 0.657, and 0.100 MPa. The yield data were explained well using the ratio of the number of excited atoms to that of ionized atoms in helium, i.e. 0.541 ± 0.005 . The scintillation yield increased with xenon partial pressure and saturated at 10% of the total pressure, while the ionization yield increased up to 1.59 times of the initial value for pure helium at 1% of the total pressure then gradually increased to 1.8 times. The helium gas with added xenon at 10% of its pressure was suggested as a medium of a neutron detector with signal readouts from both scintillation and ionization, since it increases the scintillation yield to more than 1.1 times of pure xenon and increases the ionization yield to more than increases 1.8 times of pure helium.
科研通智能强力驱动