化学
制作
粘度
工艺工程
生化工程
纳米技术
热力学
材料科学
医学
物理
工程类
病理
替代医学
作者
Anirudh Rayaprolu,S. Srivastava,Ketan Anand,Lavleen Bhati,Amit Asthana,Ch. Mohan Rao
标识
DOI:10.1016/j.aca.2018.05.036
摘要
Use of paper-based devices for affordable diagnostics is gaining interest due to unique advantages such as affordability, portability, easy disposability and inherent capillarity. As capillary transportation is an integral component of paper-based devices, low sample volume with faster measurement becomes an additional advantage. We have developed a simple, paper-based microfluidic device suitable for measuring the viscosity of Newtonian fluids as well as a few non-Newtonian fluids with sample volume as little as 12-20 μL. The results could be obtained much faster than the conventional methods. A comparative analysis of the results obtained with our paper-based viscometer and with that of the conventional Ostwald viscometer shows a correlation coefficient greater than 0.99. Apart from viscosity measurement, the paper-based devices were tested for protein denaturation and polymer molecular weight determination. Our results show that the paper-based viscometer could be a potential alternative for the conventional viscometers in the viscosity range from 0.9 cP up till ∼40 cP, with added benefits in terms of time, cost and low sample volume requirement.
科研通智能强力驱动
Strongly Powered by AbleSci AI